Presenter: D. W. Litzenberg
Presenter's Affiliation: University of Michigan
Type of Session: Plenary

Background

As set-up and interfraction errors are being reduced by in-room imaging guidance, the importance of intrafraction errors increases. Unfortunately, little data exists documenting the significance of intrafraction target movement.  The Calypso® 4D Localization System utilizes electromagnetic tracking of implanted markers to provide continuous organ monitoring. Data was collected using this investigational system to assess the relative importance of intrafraction organ movement during radiation treatment.

Materials and Methods

  • 11 patients underwent ultrasound-guided placement of 3 transponders within the apex, right base, and left base of the prostate
  • An isocenter was chosen that corresponded to the geometric center of the transponders, and patients were thereafter positioned via the electromagnetic system
  • Following initial positioning and radiographic verification, the position of the implanted transponders was monitored for 8 minutes at a rate of 10 measurements per second
  • Transponder positions were compared with previously measured data on organ position relative to skin marks, and the combined data were considered using van Herk's formula to establish margins in the following situations
    • Skin marker-based positioning
    • Pretreatment transponder positioning
    • Intratreatment continuous tracking consisting of a 3 mm threshold-based beam cutoff with repositioning and resumption of treatment

Results

  • Without consideration of intrafractional organ motion, necessary average prostate margin for skin marker-based positioning was 8.0 mm, 7.3 mm, and 10 mm in the left-right (LR), anterior-posterior (AP), and cranial-caudal (CC) directions, respectively
  • With consideration of intrafractional organ motion, necessary average prostate margin for skin marker-based positioning increased to 8.2 mm, 10.2 mm, and 12.5 mm in the LR, AP, and CC directions, respectively
  • With transponder positioning prior to each radiation fraction, necessary average prostate margin decreased to 1.8 mm, 5.8 mm, and 7.1 mm in the LR, AP, and CC directions, respectively
  • With transponder repositioning prior to treatment of each field, necessary average prostate margin again decreased to 0.4 mm, 2.3 mm, and 1.8 mm in the LR, AP, and CC directions, respectively
  • With intrafractional adjustment, necessary average prostate margin further decreased to 0.3 mm, 1.5 mm, and 1.5 mm in the LR, AP, and CC directions, respectively

Author's Conclusions

  • Interfractional, interbeam, and intrafractional adjustment allowed increasingly small margins to be employed for prostate radiation treatment
  • Significant margin reduction was accomplished by interbeam adjustment, though 2 of the 11 patients in the study with wide variation in organ motion benefited from continuous monitoring and intrafraction adjustment

Clinical/Scientific Implications

This investigational device allowed continuous non-ionizing tracking of organ motion via transponders that were amenable to placement in the path of the radiation beam without a compromise in dose distribution.  Although this technology is still early in its development, it could potentially be utilized in other treatment sites for documentation of organ motion.  Further study is needed to more fully characterize its utility and potential for clinical implementation.

English
News
ASTRO: Sexual Function Largely Retained in Prostate CA Combo Tx

Sep 18, 2014 - For patients with prostate cancer, combination beam plus brachytherapy does not compromise long-term sexual function compared with external beam radiotherapy, according to a study presented at the annual meeting of the American Society for Radiation Oncology, held from Sept. 14 to 18 in San Francisco.



I Wish You Knew

How cancer patients have changed my life

View More



Blogs and Web Chats

OncoLink Blogs give our readers a chance to react to and comment on key cancer news topics and provides a forum for OncoLink Experts and readers to share opinions and learn from each other.




OncoLink OncoPilot

Frente a un nuevo diagnóstico de cáncer o de cambiar el curso de su tratamiento actual? Deje que nuestro personal de enfermería cáncer que ayudan a pasar!

Más información